Airborne
Measurements of Formaldehyde Employing
A Tunable Diode Laser Absorption Spectrometer
The Dual Channel
Airborne Laser Spectrometer (DCALS), as configured for simultaneous
measurements of formaldehyde (CH2O)
and hydrogen peroxide (H2O2), is schematically
shown in Figure 1. During TRACE-P, only the CH2 O
channel was operational, and the discussion that follows refers
only to measurements of CH2O. Infrared (IR) radiation
at a wavelength of 3.5mm,
which is emitted
from a liquid nitrogen-cooled lead-salt diode laser, is collected
and collimated by a 90-degree off-axis parabolic mirror
(OAP). The radiation is then directed onto a second OAP mirror,
which focuses the beam through a pinhole. The radiation then expands,
hits an off-axis-elliptical (OAE) mirror and is ultimately re-focused
into the center of a multipass
astigmatic Herriott cell (Aerodyne Incorporated). The IR
beam, which trace out a Lissajous pattern in the cell, achieves
a total optical pathlength of 100 m in a 3-liter sampling volume.
After 182 reflections, the IR beam become re-entrant and exits
the cell through the entrance aperture. The
exit beam is then directed onto two different liquid-nitrogen cooled
detectors (sample detector and reference detector). As shown, the
beam in the reference arm passes through a reference cell containing
a few Torr of pure CH2O. This cell serves as wavelength
references by which to lock the laser output to the center of the
absorption feature.
Data are acquired
by repetitively sweeping the diode laser wavelength across an isolated
CH 2 O absorption
feature (2831.6417 cm -1 ) at
a frequency of 50-Hz and simultaneously applying a synchronous
50-kHz quasi-square modulation waveform to the laser tuning current.
This line is free from all spectral interferences with the exception
of a small positive interference (3.9%) from methanol. Both the
sample and wavelength reference channel is demodulated at twice
the modulation frequency employing commercial lock-in amplifiers.
The amplifier outputs are then digitized and co-averaged by computer.
The line center of the wavelength reference, which has very high
S/N, is determined on every scan and used to appropriately shift
the spectrum in memory to align the peak center before co-averaging.
Additional details regarding data acquisition and on-line processing
can be found in Fried et al. [1998a].
Sample air
is continuously drawn through the Herriott cell at flow rates around
9 standard liters minute-1 (slm) using a heated 1/2-inch
OD PFA Teflon line, which protrudes outside the aircraft boundary
layer. A winglet structure external to the aircraft allows us to
heat the inlet line to » 35 °C
to within a few cm of the inlet
entrance as well as add zero air to nearly the entire inlet. Zero
air is generated by employing a second inlet using a diaphragm
pump and a heated Pd/Al2O3 scrubber. This
scrubbing system very effectively removes CH2O without
significantly affecting the ambient water vapor concentration.
During ambient sampling, this zero air is dumped using a 3-way
Teflon valve. During background acquisition, the zero air flow
is re-routed and reintroduced back into the inlet line a few inches
from the tip. Typically the zero air flow is » 2
to 3-slm higher than the total inlet line flow. As discussed by
Fried et al. [1998a,b], background spectra are acquired approximately
twice every minute, and this approach very effectively captures
and removes optical noise as well as the effects of sample line
outgassing. In fact laboratory measurements under a variety of
relative humidities, temperatures, and sampling pressures, indicate
accurate CH2O retrieval using this approach [Wert et
al., 2001].
For
calibration purposes, CH 2 O standards
are generated using a temperature and pressure-controlled permeation
system. The CH2O permeation system contains two permeation
devices with different emission rates for calibration at two different
concentrations. The CH2O emissions rates are periodically
determined by comparing the permeation output with that generated
from a laboratory Henry’s Law calibration system. As discussed
by Fried et al. [1997a], and Gilpin et al. [1997], the calculated
output of this laboratory system has been verified by direct absorption
spectroscopy, by two different cartridge methods, and by additional
permeation devices calibrated by gravimetry. Based upon all the
collective calibrations, we estimate a total calibration uncertainty
of ± 6% at the 1s level
for both permeation devices. The calibration system output is periodically
added to the main inlet line a few inches downstream of the zero
air addition port using a separate 1/8-inch addition port. Thus we calibrate and zero nearly the
entire inlet line upstream of the sampling cell.
During airborne
operation, TDLAS measurements are acquired and stored in 5-second
increments, and 12 such measurements are obtained before a 10-second
background acquisition is acquired. Background spectra are acquired
before and after each 1-minute period using an appropriate delay
of 7-seconds (approximately 5 inlet/cell e-folding times) after
each switch. The backgrounds surrounding each ambient block are
averaged and subtracted point by point from each of the 12 ambient
spectra.
Each background-subtracted ambient spectrum thus acquired is
fit in real time to a background-subtracted
calibration spectrum acquired for each gas previously employing a
multiple linear regression approach. Typically,
the ambient-flush-background acquisition sequence is repeated for
60 minutes before a new calibration spectrum is acquired. Each
ambient measurement is corrected for drifts in laser power between
calibration acquisitions, by measuring the detector dc voltages
with and without the laser beams blocked using the lifting stage
shown in Figure 1 [Fried et al., 1998a].
Airborne replicate
measurements of CH2O
employing 1-minute averages typically yield 1s standard
deviations between 25 and 40 pptv. The
preliminary archived data are 1-minute averages. Longer averages
can be applied for improved precision. Alternatively, averages
as short as 5-seconds can be obtained for highly structured elevated
plume measurements. The final data set will present these additional
time-bases for selected interesting time periods. In addition,
the final data will be corrected for the weak methanol interference
using the onboard methanol measurements acquired by other groups.
References
Fried,
A., B. Henry, B. Wert, S. Sewell, and J.R. Drummond, Laboratory,
ground-based, and airborne tunable diode laser systems: performance
characteristics and applications in atmospheric studies, Appl.
Phys. B, 67, 317 _ 330, 1998a.
Fried,
A., B.P. Wert, B. Henry, and J.R. Drummond, Airborne tunable diode
laser measurements of trace atmospheric gases, SPIE Proc., 3285,
154 _ 162, 1998b.
Fried,
A., S. Sewell, B. Henry, B.P. Wert, T. Gilpin, and J.R. Drummond,
Tunable diode laser absorption spectrometer for ground-based measurements
of formaldehyde, J. Geophys. Res., 102, 6253 _ 6266,
1997a.
Gilpin,
T., E. Apel, A. Fried, B. Wert, J. Calvert, Z. Genfa, P. Dasgupta,
J.W. Harder, B. Heikes, B. Hopkins, H. Westberg, T. Kleindienst,
Y.-N. Lee, X. Zhou, W. Lonneman, and S. Sewell, Intercomparison
of six ambient[CH2O] measurement techniques, J. Geophys.
Res., 102, 21,161 _ 21,188,
1997.
|